P66 - time resolved Luminescence spectroscopy Beamline
The primary monochromator energy range is in between 3.7-40eV.
GENERAL SPECIFICATIONS
			
			
				
	
	| 
techniques available | 
UV- VUV spectroscopy | 
|---|---|
| 
	photon source | 
	Bending Magnet | 
| 
	length
 | 
2.3 m | 
| 
	source brilliance  | 
4.5 x 10  13 ph /s / 0.1% bw / mA    (max. 100mA) | 
| 
	polarization available | 
	linear horizontal | 
| 
	energy range  | 
3.7 eV - 40 eV | 
| 
Primary Monochromator  spectral resolution | 
resolutions are 0.02- 0.32 nm with corresponding slits 10- 2000 µm | 
| 
	spot size ON SAMPLE  | 
0.5x4 mm | 
| 
sample type | 
Solid Samples and Powders | 
OPTICS
			
			
				
	
	| type | 2 m normal incidence primary monochromator | 
|---|---|
| available gratings | 
	Al+MgF2  and  Pt  | 
| energy range | 
>3.7 eV - 40 eV | 
| resolving power | deltaE : 0.02 nm- 0.32 nm | 
| type | 
Mirror 1 and 2 | 
|---|---|
| 
Coating | 
Gold | 
| 
Energy Range | 
3.7- 40 eV | 
| 
Usage | 
M1 - Horizontal focussing | 
ENDSTATION
			
			
				
	
	| Vacuum | 
Ultra High Vacuum sample chamber with motorized sample holder for 30 samples on each side | 
|---|---|
| Temperature | 
Flow He cryostat with T range 10K - 380K, controlled by Lake Shore temperature controller | 
SECONDARY MONOCHROMATORS AND DETECTORS
			
			
				
	
	| 
UV- VIS monochromator | 
PMT (200-800nm) and CCD detector (200-1200nm) | 
|---|---|
| 
VUV-Pouey monochromator   | 
Solar Blind PMT (110-310nm)  and MSP (110-310nm) | 
| 
Absorption | 
Photodiode | 
| 
Reflectivity | 
PMT (200-800nm) | 
CONTROL/DATA ANALYSIS
			
			
				
	
	| 
Software type | 
Tango, Solis, SpectLab | 
|---|---|
| 
Data output type | 
Position of the monochromator, intensity, temperature and time  | 
| 
Data output | 
ASCII | 
SUPPORT LAB
A chemical lab with typical pellet press and most basic equipment is available.
 
						 
						 
							 
					 
					