P05 - Imaging Beamline
	 
The imaging beamline (IBL) is dedicated to micro- and nanotomography at energies between 5 keV and 50 keV. The Beamline has two end stations, a micro tomography end station for spatial resolutions down to 1 µm and FOVs between 0.9 - 7.2 mm2 and a nano tomography end station for spatial resolutions down to 100 nm and FOVs up to 50µm2.
| CONTACT | 
| 
 | 
GENERAL SPECIFICATIONS
			
			
				
	
	| techniques available | Full field tomography, micro tomography, nano tomography | 
|---|---|
| photon source | undulator | 
| period number of periods max power | 32 mm 62 - kW | 
| source brilliance | 1014 ph / s / 0.1% bw | 
| polarization available | linear horizontal | 
| energy range | 5 keV – 50 keV | 
| max flux ON SAMPLE | 1.312 ph / s / mm2 at 18 keV | 
| spot size ON SAMPLE | 
	6000 μm x6000 μm to 100 nm x100 nm        | 
| sample type | material science, geological, biological etc. | 
OPTICS
| type | Double Crystal Monochromator | Double Multilayer Monochromator | 
|---|---|---|
| Type | DESY Standard | 
Research Instruments | 
| energy range | 5 keV - 50 keV | 
5 keV - 50 keV | 
| 
	Si Crystal / ML | 
	Bragg mode 111, channel cut | 
W/B4C, Ru/C | 
| Cooling | liquid nitrogen | 
liquid nitrogen | 
ENDSTATION(S)
Microtomography
| FOV | 7.2 x 7.2 mm2 3.6 x 3.6 mm2 1.8 x 1.8 mm2 0.9 x 0.9 mm2 | 
|---|---|
| Techniques | Absorption contrast Inline phase contrast Grating based phase contast | 
| Sample environment | furnace corrosion cell load frame hanging axis | 
| Type | CCDs | 
CMOS (KIT) | 
CMOS (PCO edge 4.2) NOT YET AVAILABLE | 
|---|---|---|---|
| count rate | n.a. | 
n.a. | 
n.a. | 
| spatial resolution | down to 1 µm | 
down to 1 µm | 
down to 1µm | 
| read-out time | 3s (full frame) | 
up to 10 Hz | 
up tom 100 Hz | 
| pixel size | 
	12µm | 
6.5 µm | 
6.5 µm | 
| array size | 3k x 3k | 
5k x 4k | 
2k x 2k | 
| typical collection time | 60 min | 
down to 5 min | 
down to 1 min | 
| Software type | IDL TANGO | 
|---|---|
| Data output type | custom | 
| Data output | raw images, hdf5 | 
| Software(s) for quantitative analysis | MAVI, Volume Graphics Studio Max, Avizo | 
Nanotomography
| FOV | Contrast | 
 Spatial resolution | 
 Time resolution | 
 Field of view | 
|---|---|---|---|---|
| Tranmission X-ray microscope | Absorption contrast Zernike phase contrast Dark-field Edge measurements (absorbtion) | 
down to 40 nm (3D) | 
15 min (standard) | 
30 µm - 90 µm | 
| Holography | 
Propagation based phase contrast | 
down to < 90 nm (3D) | 
15 min (single distance) | 
30 µm - 2 mm | 
| Type | Hamamatsu C12849-111U | 
Lambda | 
Hamamatsu | 
|---|---|---|---|
| Sensor Type | sCMOS | 
Photon counting | 
sCMOS | 
| Pixel size | 6.5 µm | 
55 µm | 
5.5 µm | 
| Chip size (pixels) | 2048 x 2048 | 
516 x 516 | 
2592 x 4608 | 
| Chip size (mm²) | 13.312 x 13.312 | 
28 x 28 | 
25.344 x 13.256 | 
| Material/Scintillator | GadOx (10 µm) | 
Si (500 µm) | 
GadOx (10 µm) | 
| Max. frame rate | 30 frames /s | 
2000 frames /s | 
121 frames /s | 
| Software type | Python / TANGO | 
|---|---|
| Data output type | Images | 
| Data output | tiff | 
| Software(s) for quantitative analysis | Volume Graphics Studio Max, Avizo, Fiji, Dragonfly | 
SUPPORT LAB
- Laboratory CT
- Focussed Ion Beam (FIB) for sample preparation
- Microscopes
 
						 
						 
							 
					 
					