Unified Data Sheet

P05 - Imaging Beamline
 

The imaging beamline (IBL) is dedicated to micro- and nanotomography at energies between 5 keV and 50 keV. The Beamline has two end stations, a micro tomography end station for spatial resolutions down to 1 µm and FOVs between 0.9 - 7.2 mm2 and a nano tomography end station for spatial resolutions down to 100 nm and FOVs up to 50µm2.

CONTACT

GENERAL SPECIFICATIONS

techniques available

Full field tomography, micro tomography, nano tomography

photon source

undulator

period

number of periods

max power

32 mm

62

- kW

source brilliance 

1014 ph / s / 0.1% bw

polarization available

linear horizontal

energy range 

5 keV – 50 keV 

max flux ON SAMPLE

1.312 ph / s / mm2 at 18 keV

spot size ON SAMPLE 

6000 μm x6000 μm to 100 nm x100 nm       

sample type

material science, geological, biological etc.

 

OPTICS

type

Double Crystal Monochromator

Double Multilayer Monochromator

Type

DESY Standard

Research Instruments

energy range         

5 keV - 50 keV

5 keV - 50 keV

Si Crystal / ML

Bragg mode 111, channel cut

W/B4C, Ru/C

Cooling

liquid nitrogen

liquid nitrogen

 

ENDSTATION(S)

Microtomography

Micro Tomography

FOV

7.2 x 7.2 mm2

 3.6 x 3.6 mm2

  1.8 x 1.8 mm2

  0.9 x 0.9 mm2

Techniques

Absorption contrast

Inline phase contrast

Grating based phase contast

Sample environment 

furnace

corrosion cell

load frame

hanging axis

Detector(s)

Type 

CCDs

CMOS (KIT)

CMOS (PCO edge 4.2) NOT YET AVAILABLE

count rate 

n.a.

n.a.

n.a.

spatial resolution 

down to 1 µm

down to 1 µm

down to 1µm

read-out time 

3s (full frame)

up to 10 Hz

up tom 100 Hz

pixel size         

12µm

6.5 µm

6.5 µm

array size 

3k x 3k

5k x 4k

2k x 2k

  typical collection time

60 min

down to 5 min

down to 1 min

Control / Data analysis

Software type

IDL TANGO

Data output type

custom

Data output

raw images, hdf5

Software(s) for quantitative analysis

MAVI, Volume Graphics Studio Max, Avizo

Nanotomography

FOV

Contrast

 Spatial resolution

 Time resolution

 Field of view

Tranmission X-ray microscope

Absorption contrast

Zernike phase contrast

Dark-field

Edge measurements (absorbtion)

down to 40 nm (3D)

15 min (standard)
30 min (ROI)
< 1 min (fast)

30 µm - 90 µm

Holography

Propagation based phase contrast

down to < 90 nm (3D)

15 min (single distance)

30 µm - 2 mm
flexible

Detector(s)

Type 

Hamamatsu C12849-111U

Lambda
250k

Hamamatsu
Orca-Lightning
(available from Q4 2025)

Sensor Type 

sCMOS

Photon counting

sCMOS

Pixel size

6.5 µm

55 µm

5.5 µm

Chip size (pixels)

2048 x 2048

516 x 516

2592 x 4608

Chip size (mm²)         

13.312 x 13.312

28 x 28

25.344 x 13.256

Material/Scintillator

GadOx (10 µm)

Si (500 µm)

GadOx (10 µm)

Max. frame rate

30 frames /s

2000 frames /s

121 frames /s

Control / Data analysis

Software type

Python / TANGO

Data output type

Images

Data output

tiff

Software(s) for quantitative analysis

Volume Graphics Studio Max, Avizo, Fiji, Dragonfly

 

SUPPORT LAB

- Laboratory CT

- Focussed Ion Beam (FIB) for sample preparation

- Microscopes