P65 - EXAFS Beamline
P65 is dedicated to the majority of EXAFS experiments which do not require extremely high brilliance, photon flux or beamsizes below 1 mm2. This includes most operando experiments on catalysts, material sciences and geo- and environmental chemistry. The energy range covers 4 keV - 44 keV.
CONTACT |
|
GENERAL SPECIFICATIONS
EXAFS, XANES |
|
11 periods undulator |
|
32.8 mm 11 |
|
5x10 16 ph /s / 0.1% bw / mA (max. 100mA) |
|
linear horizontal |
|
4 keV - 44 keV |
|
deltaE/E = 1.4 x 10-4 resp. 0.6 x 10-4 |
|
2x1012 ph/s at 8.9 keV (calculated, tuned undulator) |
|
0.5 x 1 mm2 |
|
any |
OPTICS
|
Type
Water cooled Double Crystal Monochromator |
---|---|
Energy range |
4 keV - 44 keV |
Resolving power |
E/ΔE : 1x104 - 2x104
|
Available reflections |
Si111 and Si311 |
|
|
---|---|
ENDSTATION
Sample environment |
---|
DETECTOR(S)
Transmission
|
|
---|---|
Refraction
|
|
Fluorescence
|
|
Total electron yield
|
|
Grazing incidence angle
|
Type |
Ion Chambers |
---|---|
energy resolution
|
none
|
count rate
|
linear up to 1012 s-1 |
Beam Size |
given by slits |
Typical collection time |
30 s - 300 s per EXAFS scan |
Type |
Hitachi ME4 SDD
|
---|---|
Energy resolution |
|
Count rate |
2 * 105 s-1 per element |
2θ range  |
fixed at 90° |
Typical collection time per point |
0.1 s - 5 s |
CONTROL/DATA ANALYSIS
Software type
|
Sardana, Tango |
---|---|
Data output type
|
Spectra, parameters such as temp., press, etc. |
Data output |
ASCII, HDF5/NEXUS |
Software(s) for quantitative analysis
|
Larch, Demeter, FEFF |
SUPPORT LAB
A sample preparation lab equipped with a fume hood and a glove box is adjacent to the control hutch. An analytical balance and a hydraulic pellet press is available inside and outside the glove box.