P62 is dedicated to Small Angle X-ray Scattering (SAXS) experiments, which require high flux and/or energy scanning, like SAXS computed tomography and high-resolution anomalous scattering. The energy range covers 3.5 - 35.0 keV with an optimized photon flux around 12 keV. Simultaneous wide-angle X-ray scattering can be at P62.
|
CONTACT |
|---|
|
GENERAL SPECIFICATIONS
|
techniques available |
SAXS, ASAXS,GISAXS, WAXS, AWAXS |
|
photon source |
undulator U32 |
|
period / number of periods |
31.4 mm / 61 |
|
source brilliance |
1018 ph / s / 0.1% bw / mA |
|
polarization available |
linear horizontal |
|
energy range |
3.5 keV - 35.0 keV |
|
beamline energy resolution |
ΔE/E = 1.4 x 10-4 resp. 0.3 x 10-4 |
|
max flux ON SAMPLE |
1 x 1013 at 8.9 keV (calculated) |
|
spot size ON SAMPLE |
100 µm x 4 µm to 2 mm x 1 mm |
|
sample type |
any |
OPTICS
|
type |
liquid N2 cooled Double Crystal Monochromator |
|
available reflections |
Si111 and Si311 |
|
energy range |
3.5 keV - 35.0 keV |
|
resolving power |
ΔE/E : 1.4 x 10-4 .. 0.3 x 10-4 |
|
type |
Mirror 1 and 2 |
|
coating |
|
|
energy range |
3.5 keV - 35.0 keV |
|
usage |
|
|
type |
compound refractive lens |
|
material / shape |
Be / 2D |
|
radius of curvature / maximum number of lenses |
|
ENDSTATION
|
experimental table |
breadboard |
|
vacuum sample chamber |
yes |
|
beam height over table |
tbd. |
DETECTORS
|
type |
Eiger2 X 9M (in vacuum) |
|
count rate |
107 [ph/s/pixel] |
|
pixel size |
75 µm x 75 µm |
|
sensitive area |
233.1 mm x 244.7 mm |
|
array size |
3108 x 3262 = 10138296 [pixels] |
|
maximum frame rate |
230 Hz (full area) |
|
environment |
in vacuum version |
|
degree of freedom / transaltions |
3 / X-, Y-, Z-direction |
|
type |
Eiger2 X 4M-DESY |
|
count rate |
107 [ph/s/pixel] |
|
pixel size |
75 µm x 75 µm |
|
sensitive area |
155.1 mm x 162.2 mm |
|
array size |
2068 x 2162 = 4471016 [pixels] |
|
maximum frame rate |
500 Hz (full area) |
|
environment |
in air version |
|
degree of freedom / transaltions |
4 / X-, Y-, Z-direction and Tilt around Y |
SAMPLE ENVIRONMENT
|
solid sample holder |
tbd. (in vacuum and in air) |
|
liquid sample handling |
tbd. (in vacuum and in air) |
|
computed tomography setup |
tbd. (in vacuum and in air) |
|
Hexapod |
tbd. (in vacuum and in air) |
|
nanopositioning stages |
accuarcy 5nm |
|
rotation stage |
360° rotation with 100°/s |
TECHNIQUES
|
Transmission SAXS/WAXS |
yes |
|
Transmission ASAXS/AWAXS |
yes |
|
Transmission SAXS/WAXS-CT |
yes |
|
Fluorescence |
yes (detector from loan pool) |
|
Grazing incidence SAXS |
yes |
CONROL / DATA ANALYSIS
|
Software type |
Spock, Tango, Python, C++ |
|
Data output tpye |
spectra, images, parameters such as temperature, ... |
|
Data output format |
NEXUS, HDF5, ASCII |
|
Software(s) for quantitive analysis |
DPDAK, SASFIT, TomoPy, ... |