P62 is dedicated to Small Angle X-ray Scattering (SAXS) experiments, which require high flux and/or energy scanning, like SAXS computed tomography and high-resolution anomalous scattering. The energy range covers 3.5 - 35.0 keV with an optimized photon flux around 12 keV. Simultaneous wide-angle X-ray scattering can be at P62.
| 
CONTACT | 
|---|
| 
 | 
GENERAL SPECIFICATIONS
			
			
				
	
	| 
techniques available | 
SAXS, ASAXS,GISAXS, WAXS, AWAXS | 
| 
photon source | 
undulator U32 | 
| 
period / number of periods | 
31.4 mm / 61 | 
| 
source brilliance | 
1018 ph / s / 0.1% bw / mA | 
| 
polarization available | 
linear horizontal | 
| 
energy range | 
3.5 keV - 35.0 keV | 
| 
beamline energy resolution | 
ΔE/E = 1.4 x 10-4  resp. 0.3 x 10-4 | 
| 
max flux ON SAMPLE | 
1 x 1013 at 8.9 keV (calculated) | 
| 
spot size ON SAMPLE | 
100 µm x 4 µm  to 2 mm x 1 mm | 
| 
sample type | 
any | 
OPTICS
			
			
				
	
	| 
type | 
liquid N2 cooled Double Crystal Monochromator | 
| 
available reflections | 
Si111 and Si311 | 
| 
energy range | 
3.5 keV - 35.0 keV | 
| 
resolving power | 
ΔE/E : 1.4 x 10-4  .. 0.3 x 10-4 | 
| 
type | 
Mirror 1 and 2 | 
| 
coating | 
 | 
| 
energy range | 
3.5 keV - 35.0 keV | 
| 
usage | 
 | 
| 
type | 
compound refractive lens | 
| 
material / shape | 
Be / 2D | 
| 
radius of curvature / maximum number of lenses | 
 | 
ENDSTATION
			
			
				
	
	| 
experimental table | 
breadboard | 
| 
vacuum sample chamber | 
yes | 
| 
beam height over table | 
tbd. | 
DETECTORS
			
			
				
	
	| 
type  | 
Eiger2 X 9M (in vacuum) | 
| 
count rate | 
107 [ph/s/pixel] | 
| 
pixel size | 
75 µm x 75 µm | 
| 
sensitive area | 
233.1 mm x 244.7 mm | 
| 
array size | 
3108 x 3262 = 10138296 [pixels] | 
| 
maximum frame rate | 
230 Hz (full area) | 
| 
environment | 
in vacuum version | 
| 
degree of freedom / transaltions | 
3 / X-, Y-, Z-direction | 
 
| 
type  | 
Eiger2 X 4M-DESY | 
| 
count rate | 
107 [ph/s/pixel] | 
| 
pixel size | 
75 µm x 75 µm | 
| 
sensitive area | 
155.1 mm x 162.2 mm | 
| 
array size | 
2068 x 2162 = 4471016 [pixels] | 
| 
maximum frame rate | 
500 Hz (full area) | 
| 
environment | 
in air version | 
| 
degree of freedom / transaltions | 
4 / X-, Y-, Z-direction and Tilt around Y | 
SAMPLE ENVIRONMENT
			
			
				
	
	| 
solid sample holder | 
tbd. (in vacuum and in air) | 
| 
liquid sample handling | 
tbd. (in vacuum and in air) | 
| 
computed tomography setup | 
tbd. (in vacuum and in air) | 
| 
Hexapod | 
tbd. (in vacuum and in air) | 
| 
nanopositioning stages | 
accuarcy 5nm | 
| 
rotation stage | 
360° rotation with 100°/s | 
TECHNIQUES
			
			
				
	
	| 
Transmission SAXS/WAXS | 
yes | 
| 
Transmission ASAXS/AWAXS | 
yes | 
| 
Transmission SAXS/WAXS-CT | 
yes | 
| 
Fluorescence | 
yes (detector from loan pool) | 
| 
Grazing incidence SAXS | 
yes | 
CONROL / DATA ANALYSIS
			
			
				
	
	| 
Software type | 
Spock, Tango, Python, C++ | 
| 
Data output tpye | 
spectra, images, parameters such as temperature, ... | 
| 
Data output format | 
NEXUS, HDF5, ASCII | 
| 
Software(s) for quantitive analysis | 
DPDAK, SASFIT, TomoPy, ... | 
 
						 
						 
							 
					 
					