The FLUXS (FLASH Liquid-phase Ultrafast X-ray Spectroscopy) end-station is built for conducting versatile liquid-phase spectroscopic instrument for advanced solution-phase spectroscopy in the soft X-ray regime. By constructing a flexible system, this end station is able to conduct (i) X-ray absorption spectroscopy (XAS) in transmission mode, (ii) pink-beam post-sample dispersive X-ray absorption spectroscopy (dXAS), potentially in combination with beam splitting techniques for referenced detection, (iii) partial fluorescence yield (PFY) absorption spectroscopy for low-concentration samples, and (iv) resonant inelastic X-ray scattering/resonant X-ray emission spectroscopy (RIXS/RXES). The spectrometer aims for an intermediate energy resolution (E/ΔE ≤ 1000) for liquid-phase science. The compact and flexible design allows for switching between transmission and scattering schemes.
| 
								 
Target Beamlines  | 
						
						
					
				
				
					
						
							
								 
Flexible, particular at FL23, FL24, PG2 (FLASH) and P04 (PETRA III)  | 
						
						
					
				
			
|---|---|
| 
								 
Sample environment  | 
						
						
					
				
				
					
						
							
								 
liquid phase samples (30 mM)  | 
						
						
					
				
			
| 
								 
Sample Rotations  | 
						
						
					
				
				
					
						
							
								 
Vertical Jet  | 
						
						
					
				
			
| 
								 
Sample Temperature  | 
						
						
					
				
				
					
						
							
								 
Room temperature  | 
						
						
					
				
			
| 
								 
Manipulator precision  | 
						
						
					
				
				
					
						
							
								 
~1 µm  | 
						
						
					
				
			
| 
								 
Spectrometer Resolution  | 
						
						
					
				
				
					
						
							
								 
Resolving power ~2000  | 
						
						
					
				
			
| 
								 
Spectrometer Energy Region  | 
						
						
					
				
				
					
						
							
								 
Good for 200 - 800 eV  | 
						
						
					
				
			
| 
								 
Spectrometer Rotations  | 
						
						
					
				
				
					
						
							
								 
Fixed at 0° or 90°  | 
						
						
					
				
			
| 
								 
Spectrometer Polarization  | 
						
						
					
				
				
					
						
							
								 
Vertical/Horizontal  | 
						
						
					
				
			
| 
								 
Technique  | 
						
						
					
				
				
					
						
							
								 
XAS, and RIXS/RXES  |