Beamline Parameters

Source size: 142 x 5 mm2 (high-beta section)
Beam divergence: 8 x 4 urad2

2m Spectroscopy Undulator:


      Minimum gap: 9.5 mm
      Period length: 31.4 mm
      Hmax: 0.91 T
      First harmonic: 2.4 keV

Optical components

Distance from source

Properties

Double crystal
Monochromator

47 m

Si(111)
Si(311)

Phase shifter

50.5 m

Diamond and Si wave plates

First cylindrical/flat
mirror

52.3 m

Quartz / Pd-coating

Second toroidal/bent
mirror

54.6 m

Quartz / Pd-coating

High precision diffractometer in first experimental hutch

Distance from source: 76 m
Focus by mirrors: 145 x 50 um2 (FWHM) in energy range 2.7 - 24 keV
Polarization: circular and variable linear
Diffractometer: Psi-diffractometer with open Chi-circle
Complete in vacuum beampath with full polarization analysis
Sample environment: Closed-cycle cryostats

Heavy load diffractometer in second experimental hutch

Distance from source: 88 m
Focus by mirrors: 200 x 70 um2 (FWHM) in energy range 2.7 - 32 keV
Polarization: circular and variable linear
Diffractometer: Non-magnetic heavy load diffractometer
Sample environment: High-field DC magnet (14 T), He-3 insert (350 mK base temperature)

Hard X-ray Photoelectron Spectroscopy experiment in third experimental hutch

Distance from source: 94 m